Hanna GroLine Hydroponic Waterproof Pocket pH/EC/TDS/Temperature Tester HI98131
The Hanna GroLine Hydroponic Waterproof Pocket pH/EC/TDS/Temperature Tester - HI98131 is a slim, lightweight, and waterproof pocket tester that measures pH, conductivity (EC), TDS (ppm), and temperature. Simple, two-button meter operation and advanced features, including a selectable EC to TDS conversion factor (0.5 - 500 ppm and 0.7 - 700 ppm). for your essential testing.
- All measurements in a single, hassle-free tester
- Quick Cal solutions to calibrate pH and conductivity all at once with a single solution
- Perfect for hydroponics, greenhouse, and agriculture use
Simplified One-Button Calibration
One or two-point pH calibration is automatic to a pre-programmed buffer set (4.01, 7.01 and 10.01) or single point pH and EC calibration with the Hanna Quick Cal calibration solution. To perform a quick cal calibration, simply press and hold the left power/mode button while in measurement mode for three seconds and let go to enter calibration mode. Once both pH and EC are calibrated "CAL OK" is displayed and the meter returns to measurement mode.HI73127 Replaceable PH Electrode
The HI99131 TDS water tester is supplied with the HI73127 replaceable pH electrode. This electrode features an extractable junction that can be pulled 1/8 in out to clear any clogging that results in drifting and erratic readings.Features:
3 Sensors in a Single Meter
pH electrode with extractable junction, amperometric EC/TDS sensor, exposed temperature probe for faster EC/TDS readings.Selectable EC to TDS Conversion Factor
Choice of either a 0.5 or 0.7 conversion factor: 0.5 (500 CF) conversion factor is for an EC reading of 1000 μS/cm = 500 ppm; 0.7 (700 CF) conversion factor is for an EC reading of 1000 μS/cm = 700 ppm.Waterproof
Designed to withstand the humidity of a growing environment.Wide pH Range
pH range is from 0.00 to 14.00 pH.Automatic Temperature Compensation
All readings are compensated for variations in temperature Temperature displayed in °C or °F along with pH, EC or TDS reading.Stability Indicator
Meter displays a clock tag that will disappear when the reading has achieved stability.Hold Button
Freezes reading on the display to allow recording of measurement.BEPS (Battery Error Prevention System)
Meter will automatically shut off if there is not enough power to get an accurate measurement.Low Battery Indicator
Battery % level at startup. When the level is below 5%, the battery symbol on the LCD blinks to indicate a low battery condition.Auto-off
Selectable time before meter will automatically shut off to conserve battery life. Options are after 8 minutes, 60 minutes, or disabled.Specifications
pH Range | 0.00 to 14.00 pH |
pH Resolution | 0.01 pH |
pH Accuracy (@25°C/77°F) | ±0.10 pH |
pH Calibration | Automatic, one or two-point calibration (using pH 4.01, 7.01, 10.01 buffers); one point calibration using HI50036 Quick Cal pH and EC one-point calibration solution |
Temperature Range | 0.0 to 60.0 °C (32.0 to 140.0 °F) |
Temperature Resolution | 0.1 °C (0.1 °F) |
Temperature Accuracy (@25°C/77°F) | ±0.5 °C (±1.0 °F) |
EC Range | 0.00 to 6.00 mS/cm |
EC Resolution | 0.01 mS/cm |
EC Accuracy (@25°C/77°F) | ±2% F.S. |
EC Calibration | Automatic, one-point at 1.41 mS/cm or 5.00 mS/cm; one point calibration using Quick Cal pH and EC one-point calibration solution |
TDS Range | 0 to 3000 ppm (0.5 CF); 0 to 3999 ppm (0.7 CF) |
TDS Resolution | 10 ppm (mg/L) |
TDS Accuracy (@25°C/77°F) | ±2% F.S. |
TDS Conversion Factor | 0.5 (500 ppm) or 0.7 (700 ppm) |
Temperature Compensation | pH: automatic; EC: automatic, with β= 1.9%/°C |
Power Supply | 1.5V batteries (4 pcs.) |
Power Consumption | Approximately 100 hours of continuous use |
Environment | 0 to 50 °C (32 to 122 °F); RH max. 100% |
Dimensions | 163 x 40 x 26 mm (6.4 x 1.6 x 1.0 in) |
Weight | 95 g (3.4 oz) |
Ordering Information | HI98131 is supplied with HI73127 pH electrode, HI73128 electrode removal tool, batteries (4), instructions and quality certificate. |
Warranty | This meter is warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrode is warranted for a period of six months. |
Payment & Security